Option: PC-based data management software (Smart ECG Viewer)
Technical Specifications:
Input Circuit: Floating input circuit protection form
defibrillation and pacemaker
Lead: standard 12 leads
Acquisition Mode: 12 leads simultaneously
A/D converter: 12 bits
Input voltage range: +/-5mVpp
Time constant:3.2s
Frequency response: 0.05Hz-150Hz
Calibration voltage: 1mV+/-2%
Sensitivity:2.5,5,20,20,10/5,AGC (mm/mV)
Input impedance:50M 0 (10Hz)
Anti baseline drift: automatic
Filter: EMG filter 25Hz/35Hz/OFF
DFT filter 0.05/0.15/0.25 Hz
Lowpass filter 150Hz/100Hz/75Hz.
AC filter 50Hz/60Hz
CMRR: >100dB
Recording Mode: Auto/ Manual/ Rhythm
Rhythm Lead: Single lead or three leads selectable
Safety standards: IEC I/CF
Power Supply: AC: 100V-240V,50/60Hz
Battery: Built-in rechargeable Lithium-ion
battery voltage: 14.4V for about 400 copy
patient record
Recorder: Thermal paper
Record paper: Roller and Z-fold
216mm, 210mm paper width
Record Speed: 5mm/s, 10mm/s, 12.5mm/s,
25mm/s, 50mm/s +/- 3%
Dimension: 420mm x 330mm x 120mm
Display: 800 X 600 multicolor LCD
External Input/ Output: Input: >1001<0; sensitivity
10mm/V +/- 5%
Output: <1000; sensitivity
1V/mV +/- 5%
Communication interface: Ethernet, RS232, USB