Specifications
- Reconditioned & Certified with 90 Day Warranty
- Dimensions: Approximately 7 x 9 x 9 inches (18 x 23 x 23cm), excluding pole clamp protrusion and power cord storage.
-
Weight: Approximately 13lb (6.0Kg), with battery.
-
Casing: High-impact plastic.
-
Power requirements: 110-120 VAC, 50/60 Hz, 30 W
- Power Cord: Hospital-grade AC cord. 8ft long, with transparent plug and retainer plate.
-
Fuses: 0.5 A, 250 V, Slow Blowing.
-
Battery: One sealed, lead-acid rechargeable 8 V battery, internal to
device. Accessible for ease of field replacement, with color-coded
leads and polarized connector. Battery life (new batteries, full
charge, at 20 C) approximately 500 mL total volume delivered or six
hours of operation, whichever occurs first.
Special Features
DELIVERY RATE RANGE:
Micro Mode: 0.1 to 99.0 mL/hr (in 0.1 mL increments). The total of
the primary rate plus the secondary rate cannot exceed 99.9 mL/hr in
Micro Mode. Note: while using the concurrent delivery feature in the
Micro Mode, neither the primary rate nor the secondary rate can be less
than 0.5 mL/hr.
Macro Mode: 1 to 999 mL/hr (in 1 mL increments. The total of the
primary rate plus the secondary rate cannot exceed 700 mL) in 1 mL/hr
in Macro Mode.
DOSE LIMIT RANGE: Micro Mode: 0.1 to 999 mL (in 0.1 mL increments).
Macro Mode: 1 to 9999 mL (in 1 mL increments).
Operating Back Pressure: 02 to 10 psig. The maximum pressure is
user-selectable from 1 to 10 psig, via the front panel touchswitches.
Secondary Doses: For dual-channel delivery, a single dose of a
secondary fluid may be administered. For multidose delivery, 1 to 24
doses of a secondary fluid may be administered at intervals from 15
minutes to 24 hours.
OCCLUSION ALARM:
Distal: The distal occlusion alarm sounds within two pumping cycles
after the distal set tubing or set outlet fitting becomes occluded.
Proximal: The proximal occlusion alarm sounds within five pumping
cycles if the tubing proximal to the cassette becomes occluded.
Air-in-Alarm: The stopped air in distal line alarm will sound if a
bubble 100 uL or larger passes the distal air-in-line sensors. Alarm
may actually sound at detection of a bubble as small as 50 uL)
Proximal: The stopped air in proximal line alarm will sound if
approximately 1200 uL or more passes trough the proximal air-in-line
sensors.
The Lifecare 5000 Plum Infusion System is a multilane, general
purpose, volumetric infusion system designed to meet the need for
hospital-wide device standardization. By means of internal switch-settings,
a qualified bio-technician can configure the LifeCare 5000 System
to best suit the clinical needs and census changes of the hospital.
The Unit is designed for accurate, non-pulsatile administration of
single fluid regimens, dual-channel regimens, piggybacks, and multidose
regimens over a wide range of delivery rates and fluid viscosities.
It can administer a variety of medical fluids, from 5% Dextrose Injection,
USP, to enteral feeding products and blood. Primary and secondary
doses of compatible drugs can be delivered concurrently.
|
Why Shop?
- 100% Secure Ordering
- Verisign Secure Server
- We Value Your Privacy
- No Hassle Return Policy
- Great Customer Service
Customer Satisfaction:
- Honesty, integrity, persistent with service...wholehearted recommendation petehillnp– Los Angeles, CA
- Excellent product & service. Will definitely do business with Absolute Medical again, countmonk– Chicago, IL
- Honesty, integrity, persistent with service...wholehearted recommendation petehillnp– Miami, FL
- Excellent product & service. Will definitely do business with Absolute Medical again, countmonk– Houston, TX
|
|
|
|