Criticare CSI 8100 Poet Plus Vitals Signs Patient Monitor with Co2

Manufacturer: Criticare
Call 1-888-267-2966 for Pricing
Poet Plus® 8100 is the perfect choice for busy departments demanding maximum flexibility, unsurpassed ease of use, and accurate, reliable performance. Respond to every monitoring need, from standard vital signs to conscious sedation. Monitor adults, pediatrics, and neonates in virtually every department, at bedside or in transit. The Poet Plus 8100 mounts quickly and easily to walls, roll stands, and bedrails.

With the compact, lightweight Poet Plus 8100, you focus on your patients, not your monitor. The innovative user interface with simple, one-level menus and dedicated function keys ensures super-fast set-up. The big, bright display with large numerics is easily seen, even from across the room.

  • ECG, ComfortCuff™ NIBP, DOX™ SpO2, HR, Temperature, Respiration, and Nurse Call interface connection.
  • CO2, CO2/O2, IBP (2 channel), and printer

  • Easy-to-read 10.4" active TFT color display - view up to 6 waveforms simultaneously.
  • Easy to learn, easy to use menus, dedicated function keys, and simplified controls.
  • Smart battery management with rechargeable "hot-swappable" sealed lead acid batteries.
  • Lightweight design and versatile mounting options - adapts readily to department and transport requirements.
Technical Specifications

Connectors: 3- or 5-lead
Leads: User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I, II, or III (3-lead)
Gain Selection: 1/2mV, 2mV, 4mV
Frequency Response: Diagnostic: 0.05 to 100 Hz (-3db); Monitor: 0.5 to 40 Hz (-3db)
Sweep Speed: 12.5, 25, 50mm/sec
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
Pacer Detection/Rejection: Yes

Non-Invasive Blood Pressure
Technique: Oscillometric measure upon inflation
Automatic Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
Measurement Range: Adult: 30 to 300mmHg; Neonate: 0 to 150 mmHg

Resolution: 1mmHg
Transducer Accuracy: ±2mmHg or ±2% of reading, which ever is greater, over full range
STAT mode: 5 min of consecutive readings

Pulse Oximetry

Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%; <50%: unspecified
Method: Dual wavelength LED

Heart (Pulse) Rate
Source: ECG (primary); IBP, Pleth; NIBP
Range: 20 to 300bpm
Accuracy: ±1bpm or ±1% (ECG, IBP, SPO2 only)

Channels: 2
Range, Accuracy: 68° to 94.8° F (20° to 34.9° C): ±0.5%; 95° to 107.4° F (35° to 41.9° C): ±0.2%; 107.6° to 113° F (42° to 45° C): ±0.5%
Probe: YSI-400 and YSI-700 series

Rate: 6 to 150 breaths/min
Resolution: 1 breath/min
Accuracy: ±1% or ±1 breath/min, whichever is greater
Source: User selectable: Smart, CO2, ECG-TTI

Units: mmHg; Torr; %; kPa
Waveform Scale: 0 to 3.13%, 0 to 6.25%, 0 to 12.5% and 0 to 25%
Method: Non-dispersive infrared; autocalibrating
N2O Compensation: Manual; On/Off, user selectable
Range: 0 to 99mmHg; 0 to 99 Torr; 0 to 12.5%; 0 to 12.5 kPa
Resolution: 1mmHg; 1 Torr; 0.1%; 0.1 kPa
Sample Rate: Sidestream, selectable: 150ml/min or 50ml/min
Accuracy: +2mmHg, ±0.3 volume %, ±0.3 kPa, or ±2 Torr at 150ml/min and respiratory rate of ±35 breaths/min
Rise Time: 350ms at 150ml/min (10 to 90%)

Numeric Display: Inspired, expired
Method: Oxidation-reduction galvanic cell
Resolution: 1%
Range: 0 to 100%
Accuracy: ±3% volume %
Rise Time: 600ms at 150ml/min (10 to 90%)

Invasive Blood Pressure
Channels: 2
Transducer Sites: Selectable; ART, PA, LA, RA, CVP, ICP
Pressure Range: -10mmHg to 300mmHg
Zero Offset: ±150mmHg
Frequency Response: 0 to 20 Hz
Transducer Sensitivity: 5µV/V/mmHg
Zero: Menu-driven
Accuracy: ±1mmHg or ±1%, whichever is greater
Scale Range: -10 to 10, 0 to20, 0 to 30, 0 to 40, 0 to 60, 0 to 100, 0 to 150, 0 to 200, 0 to 300

Indication: Audible; visual (EN475)
Silence: Yes; 2 min or permanent

Trend Memory
Tabular: 24 hr at 30 sec intervals

Screen: 10.4" active color TFT
Resolution: 640 x 480 pixels
Waveforms: 6, maximum
Languages: English, French, German, Italian, Portuguese, Spanish

Keys: 9; membrane-activated
Rotary Knob: Push and rotate; 24 steps/turn

System Outputs
Com Ports: RS 232-compatible; digital or analog; DB9 (Com 1); mini-din analog (Com 2)
Nurse Call: Contact switch; 3.5mm stereo phone jack, standard 14mm length
Defib/Sync: BNC connector
Video Port: High density D-sub 15-pin

Type: Thermal line printer
Paper Speed: 12.5, 25, 50mm/sec

Weight: Base unit without batteries: 12 lb; 5.45kg
Size: Base unit: 13.25" (H) x 12.5" (W) x 8.75" (D); 33.65cm (H) x 31.75cm (W) x 22.22cm (D)

Power Requirements
Voltage: 100, 120, 220, 240 VAC; 50/60 Hz
Consumption: 35 W, typical
Batteries: 2; sealed lead acid; life: 2 hr, typical; recharge time: 4.5 hr
(Base unit without printer or CO2)

Operating Temperature: 59° to 95° F; 15° to 35° C
Storage Temperature: 23° to 122° F; -5° to 50° C
Operating Humidity: 15% to 90%, noncondensing

Types of Protection: Class I equipment
Degree of Protection: Type CF defib-proof
Safety: EN 60601-1

Specifications subject to change without notice.