Refurbished
- Calibrated
The CADD-Prizm VIP Model 6100 and 6101 ambulatory infusion pumps have 4 delivery modes: PCA, Continuous, TPN, and Intermittent.
Dimensions: 1.7" D x 4.1" W x 5.6" H
excluding cassette
Weight: 20 oz including 9 volt battery
and empty 100 ml reservoir
Power: 9 volt alkaline or lithium battery
OPERATING RANGE:
PCA Delivery Mode Specifications
Continuous Delivery Mode Specifications
TPN Delivery Mode Specifications
Intermittent Delivery Mode Specifications
The CADD-Prizm VIP Model 6100 and 6101 ambulatory infusion pumps have 4 delivery modes: PCA, Continuous, TPN, and Intermittent.
Dimensions: 1.7" D x 4.1" W x 5.6" H
excluding cassette
Weight: 20 oz including 9 volt battery
and empty 100 ml reservoir
Power: 9 volt alkaline or lithium battery
OPERATING RANGE:
PCA Delivery Mode Specifications
Reservoir Volume: | 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments | |
Units: | ml, mg, mcg | |
Concentration: | mg/ml, mcg/ml | |
Continuous Rate: | ![]() |
0-30 ml/hr or the mg or mcg equivalent |
Demand Dose: | 0-9.9 ml, Delivery rate is 125 ml/hr nominal | |
Demand DoseLockout: | 5 min-24 hours in the following increments: 1 min for values between 5 and 20 minutes 5 minutes between 20 min and 24 hours |
|
Max Doses Per Hour: | 1-12 doses in 1 dose increments | |
Demand Doses Given: | 0-999 | |
Demand Dose Attmepts: | 0-999 | |
Given: | 0-99999.99 in 0.01 unit increments | |
Clinician Bolus: | 0.1 ml - 20.0 ml or mg or mcg equivalent, delivery rate is 125 ml/hr nominal |
Continuous Delivery Mode Specifications
Reservoir Volume: | 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments | |
Continuous Rate: | 0.1-350 ml/hr in the following increments: 0.1 for values between 0.1-100 1 for values between 100-350 |
|
Given: | 0-99999.9 in 0.1 unit increments |
TPN Delivery Mode Specifications
Reservoir Volume: | 10.0 - 9990 or Not in Use; programmable in 1 ml increments, displayed in 0.1 increments | |
Infusion Volume: | 10-9990 ml in 10 ml increments | |
Infusion Period: | ![]() |
0 hrs 10 min - 99 hrs 40 min in 10 minute
increments |
Taper Down Period:: | 0 hrs 0 min - 99 hrs 50 min in 10 minute
increments |
|
Plateau Rate: | Calculated by pump; 10-350 ml/hr | |
KVO Rate: | Calculated by pump; 1/10 of Plateau Rate up to 5
ml/hr |
|
Given: | 0-99999 in 0.1 unit increments |
Intermittent Delivery Mode Specifications
Reservoir Volume: | 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments | |
Dose Volume: | 0.1-1000 ml in the following increments: 0.1 for values between 0.0 and 100 1 for values between 100 and 1000 |
|
Dose Duration: | ![]() |
1 min-24 hrs in the following increments: 1 min for values between 1 min and 10 min 5 min for values above 10 min Duration is limited by Dose Volume so that rate does not exceed 350 ml/hr |
Dose Cycle: | 10 min - 96 hrs in 5 minute increments | |
KVO Rate: | 0-10 ml/hr in 0.1 ml increments | |
Next Dose Start Time: | 10 min - 96 hrs X min (where X equals a 10 minute increment in the 96th hour) | |
Given: | 0-99999.9 in 0.1 unit increments |
We Also Suggest