Versatile Defibrillation and Multi-Parameter Monitoring in a Field-Proven Platform
The Physio-Control LIFEPAK® 12 Defibrillator/Monitor is a rugged, multi-parameter
defibrillator and critical care monitor designed for both pre-hospital and in-hospital use.
Built on decades of LIFEPAK innovation, it combines ADAPTIV™ biphasic defibrillation up to 360 J
with comprehensive monitoring capabilities in a single portable device, helping clinicians manage
acute cardiac and critically ill patients from first response through hospital care.
The LIFEPAK 12 supports manual defibrillation, AED mode with Shock Advisory System™,
synchronized cardioversion, and non-invasive pacing. Advanced monitoring options include
12-lead ECG acquisition and transmission, ST-segment monitoring for early ischemia
detection, Microstream® EtCO2 capnography, MASIMO SET® SpO2,
non-invasive blood pressure (NIBP), and up to two invasive pressure channels.
A large, high-resolution display and simple selector knob interface make it easy to review
waveforms and vital signs when every second counts.
Integrated data management features allow storage of CODE SUMMARY™ event records and
continuous ECG, with transmission via modem, serial, or Bluetooth® connection to LIFENET®
receiving stations for STEMI management and post-event review. Rugged construction, IPX4
splash resistance, and a wide operating temperature range support demanding EMS and
hospital environments.
This pre-owned unit delivers the same proven LIFEPAK performance and accessory
compatibility at a significantly lower cost, making it a smart choice for EMS agencies,
emergency departments, and critical care units seeking a reliable, field-tested defibrillator/monitor.
Key Features:
- ADAPTIV™ biphasic technology with escalating energy up to 360 J for difficult-to-defibrillate patients.
- Dual-mode operation with manual defibrillation and AED Shock Advisory System™ capability.
- 12-lead ECG acquisition, interpretation (GE 12SL), and ST-segment monitoring with trend review.
- Optional Microstream® EtCO2, MASIMO SET® SpO2,
NIBP, and two invasive pressure channels for advanced monitoring.
- Non-invasive pacing in demand and non-demand modes with adjustable rate and current.
- Large, high-resolution LCD/EL display with multiple waveform views and trend displays.
- Internal memory for CODE SUMMARY™, continuous ECG, trends, and vital signs; data export to
CODE-STAT™ and LIFENET® systems.
- Rugged, IPX4 splash-resistant design with extensive drop and vibration testing for EMS use.
Physio-Control LIFEPAK 12 Defibrillator/Monitor Specifications:
| Operating Modes |
Advisory (AED) mode with Shock Advisory System™;
Manual mode for ALS users; Setup, Service, and Inservice (demo) modes. |
| Waveform |
Biphasic truncated exponential with impedance compensation. |
| Energy Selection |
Manual biphasic: 2–360 J in user-selectable steps (2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100–360 J);
AED: configurable three-shock sequence, typically 150–360 J. |
| Charge Time |
Typically < 10 s to 360 J with a fully charged battery. |
| Pacing |
External demand and non-demand pacing; 40–170 ppm; 0–200 mA monophasic
truncated exponential pulse (≈20 ms). |
| ECG Monitoring |
3-, 5-, or 10-wire ECG cables (up to 12-lead acquisition);
Heart rate display 20–300 bpm; selectable ECG size and diagnostic/monitor filters. |
| SpO2 |
MASIMO SET® pulse oximetry; 1–100% saturation; adult, pediatric, and neonatal sensors;
Pulse rate 25–240 bpm (typical accuracy ±3–5 digits depending on conditions). |
| NIBP |
Oscillometric; systolic 30–245 mmHg, diastolic 12–210 mmHg, MAP 20–225 mmHg;
Accuracy typically ≤ ±5 mmHg mean, SD ≤ ±8 mmHg. |
| EtCO2 |
Microstream® capnography; 0–99 mmHg; waveform and numeric EtCO2 display;
Respiration rate 0–60 breaths/min. |
| Invasive Pressure |
Up to 2 channels; −30 to +300 mmHg; user labels (ART, PA, CVP, ICP, LAP);
Accuracy ±1 mmHg or 2% of reading (plus transducer error). |
| Display |
LCD active area approx. 140.8 × 105.6 mm (5.5 × 4.2 in) or EL 165.1 × 123.8 mm (6.5 × 4.9 in);
640 × 480 resolution; minimum 4 s ECG plus alphanumeric data; waveform sweep 25 mm/s. |
| Data Management |
Internal memory for two full patient records (CODE SUMMARY™ events plus ≈45-min continuous ECG);
Reports include 12-lead ECG, vital signs trends, and snapshots; data transfer via modem, serial, or Bluetooth. |
| Power & Batteries |
Dual-battery operation with FASTPAK®, FASTPAK 2, LIFEPAK NiCd, or SLA batteries;
Optional AC power adapter with on-board battery charging. Typical run-times vary by battery type and use
(monitoring, pacing, or high-energy shocks). |
| Physical Characteristics |
Weight (unit + QUIK-COMBO® cable, no batteries): approx. 6.6 kg (14.5 lb);
Height 31.7 cm (12.5 in); Width 39.6 cm (15.6 in); Depth 23.1 cm (9.1 in). |
| Environmental |
Operating temperature 0–50 °C (32–122 °F); storage −20–60 °C (−4–140 °F);
Relative humidity 5–95% non-condensing; splash-proof to IPX4; tested to 18 in drop and MIL-STD-810E vibration. |
| Typical Applications |
EMS transport, fire/EMS vehicles, emergency departments, cath labs, and critical care units;
Adult and pediatric defibrillation, cardioversion, pacing, and advanced physiologic monitoring. |