Invivo MDE Escort M8 Patient Monitor

Invivo MDE Escort M8 Patient Monitor

Vendor: Invivo
$695.00 Buy Now
Please, enter quantity for at least one product
Quantity unavailable. Try selecting a smaller quantity.
More Product Pictures

Electrosurgery. The Escort M8 Vital Signs Monitor is suitable for use in the presence of electrosurgical (ESU) equipment. The following precautions should be taken:

  • To minimize the risk of patient burns, only use ESU equipment that monitors the impedance of the ESU return wires.
  • Users should be properly trained in the operation of the ESU equipment.
  • Keep patient applied cables (e.g. ECG lead wires) off of earth ground and away from the ESU knife and return wires.
  • Only use Invivo Research recommended accessories.
  • While cutting, use the SpO2 parameter instead of the ECG parameter to determine heart rate.

Alarm Monitoring. Failure to respond to alarms that are annunciated by the Escort M8 Vital Signs Monitor will cause a lapse in patient monitoring. Always respond promptly to alarms.

Flammable Anesthetics. An explosion hazard exists if the monitor is used in the presence of flammable anesthetics.

Anesthesia Patients. Constant attention by a qualified individual is needed whenever a patient is under anesthesia or connected to a ventilator.

Device Interconnections. Through its Network Connector port, the Escort M8 Vital Signs Monitor can be connected to external devices. The following precautions should be taken:

  • The Network Connector cable should not be applied to the patient.
  • Connected devices should be located outside of the patient vicinity (greater than 1.5 meters) if they do no comply with IEC 60601-1.
  • The Escort M8 Vital Signs Monitor should not be connected to devices that are not described in this manual.
  • The over-all system leakage current should be tested and should comply with IEC 60601-1-1.

Do not operate the Escort M8 Vital Signs Monitor near high frequency emissions (e.g. microwaves).

Single use devices should not be reused.

The accuracy of the measurements can be affected by the position of the patient, the patient’s physiological condition, and other factors. Always consult a physician for interpretation of measurements made by this monitor.

If any system failure occurs (e.g. an unexplained continuous audible alarm) remove the monitor from use, and refer it to qualified service personnel. Perform operational checkout before each use. If monitor fails to function properly, refer to qualified service personnel.

For safe and accurate operation, use only recommended Invivo Research patient cable, lead wires, cuffs, hoses, sensors, tubing, etc. A listing of these can be found in the Accessory Listing within this manual, or by contacting Invivo Research directly.

The system may not conform to all performance specifications if stored or used outside the environmental specifications identified in Section 12.

For proper equipment maintenance, perform the service procedures at the recommended intervals as described in the monitor’s service manual.

Indications for Use

The Escort M8 Vital Signs Monitor is indicated for use in adult and pediatric patient populations in an environment where patient care is provided by healthcare professionals (e.g. Physician, Nurse, Technician) when:
1. The professional determines that a device is required to measure any or all of the following patient parameters:

  • Heart Rate
  • Respiration Rate
  • Temperature
  • Systolic, Mean and Diastolic Blood Pressure
  • Functional Oxygen Saturation of Arterial Hemoglobin (SpO2)
  • Pulse Rate

And,

2. The professional determines that a device is required to issue visible and audible alarms when any or all of the following parameters exceed preset limits:

  • Heart Rate
  • Respiration Rate
  • Temperature
  • Systolic, Mean and Diastolic Blood Pressure
  • Functional Oxygen Saturation of Arterial Hemoglobin (SpO2)
  • Pulse Rate